How should the exhaust gas be handled after ozone sterilization of medical devices?

After medical devices are sterilized with ozone, unreacted ozone may remain in the sterilization chamber, exhaust piping, and related equipment units. Controlled exhaust and appropriate tail gas treatment are needed to reduce its concentration. Because ozone is strongly oxidizing and irritating to the respiratory tract, uncontrolled exhaust can increase the risk of ozone exposure in the workplace. For equipment that uses an ozone sterilization process, a suitable tail gas treatment system can be configured based on the exhaust volume per cycle, instantaneous flow rate, ozone concentration, gas composition, and operating cycle. Ozone decomposition catalysts can promote the conversion of ozone to oxygen and are one of the optional treatment materials. However, whether the system can operate stably depends not only on catalyst activity but also closely on exhaust control, airflow distribution, reaction conditions, detection methods, and equipment interlocks.

I. What role does ozone play in the medical device sterilization process?

Ozone is a gas with strong oxidizing capability that can destroy microbial cell structures and biological activity through oxidation, and is therefore used in sterilization processes for certain medical devices. Compared with liquid disinfection methods, gaseous ozone has a certain diffusion capability and can penetrate device surfaces and some complex structural areas, but the actual sterilization effect still depends on process conditions such as ozone concentration, exposure time, temperature, humidity, and gas distribution.

In a typical ozone sterilization device, ozone is generated by a dedicated generator and enters the sterilization chamber according to a preset program. The equipment controls ozone supply, chamber environment, and exposure time so that the sterilization process meets predetermined requirements. For devices with small channels, complex structures, or combinations of different materials, the actual ozone contact conditions and material compatibility also need to be considered.

It should be noted that ozone sterilization is not suitable for all medical devices. The strong oxidizing properties of ozone may affect some rubbers, polymers, and other sensitive materials, and device manufacturers need to determine suitability based on material tolerance and sterilization process validation results. ISO 14937:2009 specifies general requirements for characterization of the sterilizing agent, process development, validation, and routine control in medical device sterilization and can serve as a reference framework for the development and validation of a specific sterilization process, but it cannot replace validation work for a specific process.

II. Where does residual ozone come from after sterilization?

The end of the sterilization cycle does not mean that all ozone inside the equipment has decomposed. Some ozone may still be present in the chamber, piping, and related gas spaces and must be removed through equipment exhaust, purging, or other specified procedures.

1. Unreacted ozone in the sterilization chamber

During sterilization, ozone contacts microorganisms, device surfaces, and other oxidizable substances in the chamber. Because reaction extent and gas distribution vary, some ozone may not participate in reactions and may remain in the chamber space. The residual amount is related to factors such as initial ozone concentration, sterilization time, chamber structure, load, and surface reactions.

2. Residual ozone in exhaust piping and equipment components

After sterilization, gases in the chamber usually need to be discharged through valves, piping, and connectors. Ozone-containing gas retained in the piping also enters the tail gas treatment device during exhaust. If the piping is long, has branches, or has local gas stagnation, detecting only the ozone concentration inside the chamber may not fully reflect the final ozone load discharged.

3. Intermittent exhaust creates instantaneous loads

Medical device ozone sterilization equipment usually operates in batches or cycles, and the exhaust process is intermittent. The equipment may discharge a certain volume of ozone-containing gas in a relatively short period, causing the treatment device to experience short-term flow peaks and concentration fluctuations.

Therefore, the tail gas treatment system cannot be designed solely based on the average exhaust volume of the equipment. It is also necessary to understand the exhaust volume per cycle, exhaust duration, exhaust frequency, and whether multiple units may exhaust simultaneously.

III. Why does tail gas from medical device ozone sterilization require dedicated treatment?

Ozone is irritating and may adversely affect the eyes and respiratory system after inhalation. The occupational exposure limits for ozone published by NIOSH and OSHA in the United States are both 0.1 ppm, but the definitions and applicable systems of the two limits differ. Actual projects should follow local regulations, occupational hygiene requirements, and applicable standards, and a limit from one country cannot be directly treated as a unified emission standard for all regions.

If residual ozone is removed only by natural decomposition, the actual effect will be affected by factors such as temperature, humidity, gas flow, spatial conditions, and residence time. It is difficult for natural decomposition to directly ensure that the target concentration is reached within the specified time after each sterilization cycle. Therefore, for equipment with personnel activity and periodic exhaust, a controlled exhaust and treatment path should be established based on site conditions.

The tail gas treatment system needs to address three issues:

  1. Control gas collection: Route ozone-containing gas discharged from the equipment into designated piping to reduce fugitive dispersion in the work area.
  2. Reduce ozone concentration: Select appropriate decomposition or other treatment methods based on exhaust load and target requirements.
  3. Confirm treatment effectiveness: Use reasonable detection points and operating records to determine whether the treatment device can operate stably across different sterilization cycles.

It is important to distinguish that validation of the sterilization process itself and performance validation of the tail gas treatment system are different tasks. A tail gas treatment device cannot replace sterilization effectiveness validation, nor should it change the exhaust program of the original sterilization equipment without evaluation.

IV. How are ozone decomposition catalysts used in medical device sterilization tail gas treatment?

Ozone decomposition catalysts can promote the decomposition of ozone on the material surface, converting ozone into oxygen. The basic reaction is:

2O₃ → 3O₂

During catalysis, ozone molecules contact the catalyst surface and decompose under the action of active sites. The catalyst lowers the energy barrier required for the reaction, allowing ozone to convert faster under suitable conditions. Compared with relying solely on natural decomposition, catalytic treatment can provide a relatively controllable reaction step for tail gas purification.

1. Typical tail gas treatment process

For equipment using an external tail gas treatment device, the following treatment process can be established based on the actual structure:

  1. Equipment exhaust: After the sterilization cycle ends, the equipment discharges ozone-containing gas from the chamber according to a validated program.
  2. Enclosed collection: Transport the gas to the treatment device through exhaust piping to reduce piping leakage and fugitive emissions.
  3. Flow regulation: Based on the exhaust volume per cycle and instantaneous flow rate, provide buffering or regulation as needed to prevent short-term loads from exceeding the treatment device capacity.
  4. Catalytic decomposition: Pass the ozone-containing gas through a reactor packed with ozone decomposition catalyst so that ozone decomposes on the catalyst surface.
  5. Treatment effect detection: Set appropriate detection points at the reactor inlet and outlet to confirm changes in ozone concentration and determine the subsequent discharge method according to site requirements.

Actual equipment may also use an integrated catalytic treatment structure. For example, some ozone sterilization devices decompose residual ozone through an internal catalytic unit after the sterilization cycle ends. Whether to use internal or external treatment depends on the equipment design, manufacturer requirements, and actual exhaust path.

2. Applicability boundaries of catalysts

Ozone decomposition catalysts primarily target gaseous ozone and do not mean that volatile organic compounds, particulate matter, oil mist, or other pollutants in the tail gas can be removed at the same time. If these components are present in the tail gas, their types and concentrations need to be analyzed to determine whether they will affect catalyst activity, cause surface contamination, or cause bed blockage, and a pretreatment unit should be configured when necessary.

Different catalysts may differ in temperature, humidity, space velocity, and impurity tolerance. Especially under high-humidity conditions, water molecules may compete with ozone for active sites on the catalyst surface and affect the ozone decomposition process. Therefore, applicability cannot be judged solely based on the catalyst name or theoretical reaction principle; testing should be carried out under actual gas conditions.

V. How should the key design parameters of a tail gas treatment system be evaluated?

System design should start from actual exhaust conditions rather than estimating solely based on sterilization chamber volume or catalyst loading.

1. Evaluate instantaneous flow based on exhaust volume per cycle

If the exhaust volume per cycle and exhaust duration are known, the average exhaust flow rate can be initially estimated:

Q = V / t

Where:

  • Q: average exhaust flow rate;
  • V: exhaust volume per cycle;
  • t: exhaust duration.

For example, if equipment exhaust is not uniform, flow peaks during the exhaust process should also be identified through actual measurement or equipment operating records. When designing the treatment device, attention should be paid to the actual peak and its duration, rather than using only the average flow rate over the entire sterilization cycle.

For cases where multiple units share one treatment system, the equipment operating sequence also needs to be analyzed. If multiple units may exhaust simultaneously, the simultaneous operating load should be included in the design; if exhaust times are staggered through control programs, the reliability of the control logic needs to be confirmed, and load changes under abnormal conditions should be considered.

2. Estimate ozone treatment load based on flow and concentration

The ozone treatment load per unit time can be estimated from gas flow rate and ozone concentration:

M = Q × C

Where:

  • M: mass of ozone entering the treatment device per unit time;
  • Q: volumetric gas flow rate;
  • C: inlet ozone mass concentration.

When Q is in cubic meters per hour and C is in milligrams per cubic meter, the unit of M is milligrams per hour. If ozone concentration is expressed in ppm, unit conversion must be performed based on temperature, pressure, and gas conditions; the ppm value cannot be directly multiplied by volumetric flow rate and used as a mass load.

This estimate helps determine differences in ozone load among different sterilization cycles, but it cannot be used alone to determine catalyst loading. Catalyst usage is also related to reaction temperature, space velocity, contact time, catalyst performance, and actual airflow distribution.

3. Pay attention to space velocity and effective contact time

Space velocity is usually used to describe the volumetric gas flow passing through a unit volume of catalyst bed per unit time. Its calculation requires clarification of the reference conditions for gas flow measurement and the catalyst bed volume.

When other conditions are similar, an increase in space velocity means that a unit volume of catalyst must treat more gas, and the effective contact time between the gas and the catalyst usually becomes shorter. If the space velocity is too high, ozone may leave the bed before it is fully decomposed.

However, space velocity is not the only indicator for judging catalyst performance. Actual treatment effectiveness also depends on catalyst activity, temperature and humidity, ozone concentration, bed structure, and gas distribution. Engineering design should determine a reasonable range based on test data for the catalyst under target conditions.

4. Temperature, humidity, and gas impurities

Temperature and humidity affect the adsorption and decomposition of ozone on the catalyst surface. Under high-humidity conditions, water molecules may occupy some active sites and affect effective contact between ozone and the catalyst. The degree of this effect differs among material systems and needs to be determined through testing under specific conditions.

If the tail gas contains oil mist, particulate matter, or other impurities, their effects on the catalyst surface and bed pressure drop should also be evaluated. For components that may cause catalyst poisoning or clogging, pretreatment requirements and maintenance plans should be defined during the design stage.

5. Reactor structure and airflow distribution

Even if the catalyst itself has good ozone decomposition activity, uneven airflow distribution in the reactor may lead to problems such as excessively high gas velocity in some areas, insufficient utilization of local bed volume, or airflow short-circuiting.

Reactor design needs to comprehensively consider inlet gas distribution, bed thickness, catalyst loading method, support structure, and system pressure drop. Proper airflow organization can enable gas to pass through the catalyst bed more uniformly and reduce the risk of insufficient local treatment.

VI. How can the actual treatment capacity of an ozone catalytic decomposition system be verified?

Whether the system achieves the expected treatment effect needs to be confirmed through representative operating tests and cannot be judged solely based on catalyst sample parameters or theoretical reaction efficiency.

1. Detect inlet and outlet ozone concentrations

Set detection points at the catalytic reactor inlet and outlet to measure the ozone concentrations entering and leaving the reactor. Using the two concentration data, the ozone decomposition efficiency under specific test conditions can be calculated:

η = (Cin − Cout) / Cin × 100%

Where:

  • η: ozone decomposition efficiency;
  • Cin: reactor inlet ozone concentration;
  • Cout: reactor outlet ozone concentration.

For example, when the inlet concentration is 10 mg/m³ and the outlet concentration is 1 mg/m³, the concentration reduction ratio calculated by the above formula is 90%. This is only a calculation example and does not represent the actual performance of any specific catalyst or equipment.

It should be noted that the concentration reduction ratio is not equal to stable treatment capacity under all operating conditions. If the inlet flow rate, concentration, temperature, or humidity changes, the test results may also differ. Therefore, the flow rate, temperature, humidity, exhaust stage, and equipment operating status during the test should also be recorded.

2. Select representative test conditions

The exhaust of medical device ozone sterilization equipment is periodic, and a single test may not reflect the complete operating situation. Testing should cover representative sterilization cycles, with particular attention to:

  • Inlet and outlet ozone concentrations during the normal exhaust stage;
  • Flow rate and concentration changes during short-term concentrated exhaust;
  • Treatment performance when multiple units operate simultaneously or continuously;
  • Outlet concentration changes under different temperature and humidity conditions;
  • Treatment effectiveness and pressure drop changes after continuous operation for a period of time.

If the equipment has different exhaust programs, the treatment load corresponding to each program should be confirmed separately to avoid using only one lower-load condition to represent all operating states.

3. Ensure comparability of detection data

Ozone concentration detection results are affected by factors such as instrument range, detection principle, sampling line, sampling location, and environmental conditions. Inlet and outlet detection should use instruments suitable for the target concentration range and should be calibrated or performance-checked according to equipment requirements.

Sampling points should be located so that they can represent the actual gas conditions at the reactor inlet and outlet as much as possible, and should be avoided in locations where gas mixing is insufficient or where dilution by outside air is likely. For intermittent exhaust processes with rapid concentration changes, instrument response time and data logging frequency should also be considered to avoid missing short-term peaks.

4. Establish an abnormal condition diagnosis mechanism

During system operation, if the outlet ozone concentration rises, it should not be directly concluded that the catalyst has failed; instead, troubleshooting should be carried out in combination with inlet conditions and equipment status.

Abnormal symptom Possible cause Recommended checks
Elevated outlet ozone concentration Increased inlet load, excessive space velocity, or decreased catalyst activity Check inlet concentration, flow rate, and catalyst condition
Increased system pressure drop Dust accumulation in the bed, clogging, or catalyst powdering Check the bed, filter components, and gas impurities
Fluctuating treatment performance Intermittent exhaust, uneven airflow distribution, or temperature/humidity changes Check exhaust sequence, flow rate, and reaction conditions
Local gas leakage Abnormal piping connections or sealing components Check valves, flanges, and piping sealing

For the catalyst bed, the loading condition, mechanical integrity, and pressure drop changes should also be inspected regularly. If obvious powdering, damage, or contamination occurs, whether maintenance or replacement is needed should be determined based on treatment effectiveness and operating records.

VII. What should be considered when selecting a tail gas treatment system for medical device ozone sterilization?

Tail gas treatment system selection should be based on the actual equipment conditions and validation objectives. The following steps are recommended:

  1. Collect equipment information: Clarify sterilization chamber volume, ozone generation rate, exhaust program, exhaust volume per cycle, exhaust time, and operating frequency.
  2. Analyze tail gas conditions: Obtain inlet ozone concentration, instantaneous flow rate, temperature, humidity, and other gas components, and identify impurities that may affect the catalyst.
  3. Confirm treatment objectives: Based on applicable occupational hygiene, environmental emission, and equipment safety requirements, clarify the treatment device outlet concentration or other acceptance criteria.
  4. Evaluate catalyst applicability: Conduct sample testing or small-scale trials based on target temperature and humidity, ozone load, space velocity, and impurity conditions.
  5. Determine system structure: Comprehensively consider independent treatment for a single unit or centralized treatment for multiple units, airflow distribution, bed pressure drop, piping sealing, and monitoring points.
  6. Conduct on-site verification: Detect inlet and outlet ozone concentrations during representative operating cycles, record flow rate, temperature and humidity, pressure drop, and abnormal conditions, and confirm whether the system can meet predetermined requirements.

For equipment manufacturers and medical device manufacturers, the interface relationship between the tail gas treatment device and the sterilization equipment control system should also be clarified. For example, when the catalytic device fails, the fan is abnormal, or the outlet ozone concentration does not meet the set requirements, how the equipment should alarm, pause exhaust, or take other safety measures needs to be defined during the system design stage. The specific interlock strategy should be determined by the equipment designer based on risk assessment and applicable regulations, and operational safety cannot be ensured by the catalyst alone.

Conclusion

Tail gas treatment after medical device ozone sterilization requires simultaneous consideration of residual ozone sources, equipment exhaust characteristics, catalytic reaction conditions, and on-site safety management. Ozone decomposition catalysts can promote the conversion of ozone to oxygen, but their actual treatment capacity depends on multiple factors such as inlet concentration, instantaneous flow rate, temperature and humidity, space velocity, airflow distribution, and catalyst condition.

For intermittently operating sterilization equipment, particular attention should be paid to the exhaust load per cycle and short-term flow peaks, and sizing should not be based solely on average flow rate. After the system is put into operation, treatment effectiveness should also be verified through inlet and outlet ozone concentration detection, pressure drop monitoring, and representative cycle testing, and an abnormal diagnosis and maintenance mechanism should be established. Only by combining catalyst performance, equipment exhaust control, reactor design, and operational monitoring can controlled treatment of ozone sterilization tail gas be achieved more reliably.